Thermocouple Equivalent Temperature
Accuracies (Readout Only)
Type 'B'
±0.85°C
from 600°C
to 800°C
±0.68°C
from 800°C
to 1000°C
±0.57°C
from 1000°C
to 1800°C
Type 'C'
±0.32°C
from 100°C
to 550°C
±0.71°C
from 550°C
to 2300°C
Type 'E'
±0.52°C
from -200°C
to 0°C
±0.22°C
from 0°C
to 950°C
Type 'J'
±0.52°C
from -200°C
to 0°C
±0.23°C
from 0°C
to 1200°C
Type 'K'
±0.61°C
from -200°C
to 0°C
±0.24°C
from 0°C
to 1370°C
Type 'L'
±0.36°C
from -200°C
to 0°C
±0.23°C
from 0°C
to 1370°C
Type 'M'
±0.26°C
from -20°C
to 0°C
±0.25°C
from 0°C
to 400°C
±0.22°C
from 400°C
to 1400°C
Type 'N'
±0.72°C
from -200°C
to 0°C
±0.28°C
from 0°C
to 1300°C
Type 'R'
±1.09°C
from -20°C
to 0°C
±0.97°C
from 0°C
to 500°C
±0.49°C
from 500°C
to 1750°C
Type 'S'
±1.05°C
from -20°C
to 0°C
±0.95°C
from 0°C
to 500°C
±0.56°C
from 500°C
to 1750°C
Type 'T'
±0.60°C
from -200°C
to 0°C
±0.25°C
from 0°C
to 400°C
Type 'U'
±0.54°C
from -200°C
to 0°C
±0.24°C
from 0°C
to 400°C
Note: Accuracies are based on
internal Reference Junction Compensation. Refer to Technical manual