Test Voltage 500 V DC |
Range: 200 kW to 10 GW
Accuracy: 5%
Range: 10 GW to 100 GW
Accuracy: 20% |
|
Test Voltage 1000 V DC |
Range: 200 kW to 20 GW
Accuracy: 5%
Range: 20 GW to 200 GW
Accuracy: 20 |
|
Test Voltage 2500 V DC |
Range: 200 kW to 50 GW
Accuracy: 5%
Range: 50 GW to 500 GW
Accuracy: 20% |
|
Test Voltage 5000 V DC |
Range: 200 kW to 100 GW
Accuracy: 5%
Range: 100 GW to 1 TW
Accuracy: 20% |
|
Short circuit current |
> 1 mA
and < 2 mA |
|
Bar graph |
Range: 0 to 1 TW |
|
Insulation test voltage |
Accuracy: 0% to 10% at 1 mA load current |
|
Induced AC mains current rejection |
Max: 2 mA |
|
Charging rate for capacitive load |
5 s/µF |
|
Leakage current |
Range: 1 nA to 2 mA
Accuracy: ±(5% + 2 nA) |
|
Capacitance measurements |
Range: 0.01 µF to 15.00 µF
Accuracy: ±(15% rdg + 0.03 µF) |
|
Live circuit indicator |
Range: 30 V to 600 V (AC/DC, 50/60 Hz)
Accuracy: ±(5% + 2V) |
|
Timer |
0 - 99
minutes
Settable: 1 minute increments
Indication: 1 second resolution |
|
Ramp |
0% to
100% of selected test voltage, or until breakdown |