Model 1745 Three-Phase Power
Quality Logger - Specifications |
|
|
|
| |
| |
 |
| |
Intrinsic Error |
Refers to the reference
conditions and is guaranteed for two years. |
| Warranty |
Two years |
| Recalibration Interval |
Two years recommended |
Quality System |
Developed, designed, and
manufactured according to DIN ISO 9001. |
Reference Conditions
|
| |
23°C
2K; 74°F ±2K, Vm=230V ±10%, 50 Hz ±0.1 Hz or 60 Hz ±0.1 Hz |
| Phase
Sequence: |
L1, L2, L3 |
| Interval length: |
10 minutes |
| Wye
connection: |
(l1, L2,
L3 to N) |
| Power
supply: |
88 volts
to 265 volts AC |
|
| Environmental Condition
Ranges |
| Working
Temperature: |
-10°C
to 55°C; 14°F
to 131°F |
| Operating
Temperature: |
0°C
to 35°C; 32°F
to 95°F |
| Storage
Temperature: |
-20°C
to 60°C; -4°F
to 140°F |
| Reference
Temperature: |
23°C
±2K; 74°F ±2K |
|
Relative Humidity |
| Model 1743: |
Class C2 acc. IEC 60654-1 |
| Model 1744: |
Class C2 acc. IEC 60654-1 |
| Model 1745: |
Class B2 acc. IEC 60654-1 |
| Housing |
Robust, fully insulated
housing and accessories |
| Protection
Type |
| Model 1743: |
IP65 as per EN 60529 |
| Model 1744: |
IP65 as per EN 60529 |
| Model 1745: |
IP50 as per EN 60529 |
Safety |
IEC/EN 61010-1 600V CAT III,
300V CAT IV, pollution degree 2, double insulated |
| Type Test Voltage |
5.2 kV rms, 50 Hz/60Hz, 5
sec |
EMC |
| Emission: |
IEC/EN
61326-1, EN 55022 |
| Immunity: |
IEC/EN
61326-1 |
|
|
| |
| |
 |
| |
|
Input Voltage |
|
|
Input
range VIP-N: |
maximum
480 V AC |
|
Input
range VIP-P: |
maximum
830 V AC |
|
Max.
overload voltage: |
1.2 VI |
|
Input
range selection: |
by job
programming |
|
Connections: |
P-P or
P-N, 1- or 3-phase |
|
Nominal voltage VN: |
less than
or equal to 999 kV with PTs and ratio |
|
Input
resistance: |
approx.
820 KΩ per channel Lx-N
single-phase (L1 or A, L2 or B, L3 or C connected) approx.
300 KΩ |
|
Intrinsic uncertainty: |
0.1% of VI |
|
Voltage transformer: |
ratio:
<999 kV/V |
|
Ration
selection: |
by job
programming |
|
|
Current Input with Flexi Set |
|
|
Input
ranges II L1 or A, L2 or B, L3 or CN: |
15A /150A
/ 1500A / 3000A AC |
|
Measuring Range: |
0.75A to
3000A AC |
|
Intrinsic uncertainty: |
<2% of II |
|
Position influence: |
maximum
±2% of measured value - for distance conductor to
measurement head >30mm |
|
Stray
field influence: |
< ±2A AC
for Iext=500A AC and distance to measuring head >200mm |
|
Temperature coefficient: |
<0.05%/k |
|
Current transformer: |
ratio less
than or equal to 999 kA/II |
|
Ratio
selection: |
by job
programming |
|
Connection: |
3-phase,
3-phase +N, 2-phase L1 or A and L3 or C (2
Wire-meter-method) 7 pole connector |
|
|
Current Input for Clamp |
|
|
Input
ranges II L1 or A, L2 or B, L3 or C, N: |
0.5V
nominal (for II) 1.4V peak |
|
Intrinsic uncertainty: |
<3% of II |
|
Maximum overload: |
10V AC |
|
Input
resistance: |
Approx.
8.2KΩ |
|
Current transformer: |
ratioΩ
999kA/II |
|
| Power Systems |
Delta, 2-Element
Delta, Wye, Single-Phase, Split Single-Phase |
|
| |
| |
 |
| |
|
Power Supply |
|
Functional range: |
|
Power
consumption: |
|
Ride
through model 1743: |
|
Ride
through model 1744: |
|
Ride
Through model 1745: |
|
Fuse:
|
|
|
88V to
660V absolute, 50Hz/60Hz 100V to 350V DC internal fuse 630mA T |
|
5 watts |
|
3 second
capacitor |
|
3 second
capacitor |
|
Internal
battery for typical > 5 hours ride through with intelligent
power management |
|
Power
supply fuse can be replaced in service facility only. Supply can be connected in parallel to measuring inputs (up
to 660V) |
|
Display Indicators
|
LED's for status
and voltage levels Fluke model 1745: LC-display with backlight for voltage, current,
active power, phase sequence |
| Memory |
|
Intervals:
|
|
Events: |
|
Memory
model: |
|
|
Capacity 8
MB Flash-EPROM |
|
A function
>12000 intervals for >85 days with 10 minute intervals. P function >30000 intervals for >212 days with 10 minute
intervals. |
|
>13000 |
|
linear,
circular |
|
Interface |
RS232, 9600
...115000 baud, automatic baud rate selection, 3-wire
communication |
| Measurement |
|
A/D
converter: |
|
Anti-aliasing filter: |
|
Frequency response: |
|
Interval length: |
|
Averaging time for Min/Max values:
|
|
Time base |
|
Resolution: |
|
Deviation: |
|
| |
|
16 bit,
sample rate: 10.24 kHz |
|
FIR-Filter, fc=4.9 kHz |
|
uncertainty <1 % of Vm for 40 Hz to 2500 Hz |
|
1, 3, 5,
10, 30s, 1, 5,10, 15, 60 minutes |
|
1/2, 1
mains period, 200 ms, 1, 3, 5 seconds |
|
|
|
10ms (at
50Hz) |
|
2s/day at
23°C (74°F)
±2K |
|
|
Dimensions |
|
Model
1743: |
|
Model
1744: |
|
Model
1745: |
|
| |
|
170mm (H)
x 125mm (W) x 55mm (D) 69" (H) x 51" (W) x 22" (D) |
|
170mm (H)
x 125mm (W) x 55mm (D) 69" (H) x 51" (W) x 22" (D) |
|
282mm (H)
x 216mm (W) x 74mm (D) 115" (H) x 88" (W) x 33" (D) |
|
|
Weight |
|
Model
1743: |
|
Model
1744: |
|
Model
1745 |
|
|
|
|