| Infrared temperature
range |
-40°C to
800°C (-40°F
to 1472°F) |
| Infrared accuracy |
<0°C
(32°F): ±(1.0°C+
0.1°/1°C)/±(2.0°F+
0.1°/1°F)
>0°C (32°F):
±1% or ±1.0°C
(±2.0°F)
whichever is greater |
| Display resolution |
0.1°C or 0.1°F |
| Infrared spectral
response |
8µm to 14µm |
| Infrared response time |
<500msec |
Thermocouple Type-K
input
temperature range |
-270°C to
1372°C (-454°F
to 2501°F) |
| Thermocouple Type-K
accuracy |
-270°C
to -40°C: ±(1°C+
0.2°/1°C)
-454°F to -40°F:
±(2.0°F+ 0.2°/1°F)
-40°C to 1372°C:
±1% or 1°C
whichever is greater
-40°F to 2501°F:
±1% or 2°F whichever is
greater |
Distance to measurement
spot size (D:S) |
50:1 |
| Laser sighting |
Single-point laser <1mw output Class 2
(II) operation, 630nm to 670 nm |
| Minimum spot size |
19mm (0.75 in) |
| Emissivity adjustment |
By built-in table of common materials
or digitally adjustable from 0.10 to 1.00 by 0.01 steps. |
Data storage with
Date/Time stamp |
99 points |
| PC interface and cable |
USB 2.0 with FlukeView®
Forms software |
| High/Low alarms |
Audible and two-color visual |
| Min/Max/Avg/Dif
functions |
Yes |
| Display |
Dot matrix 98 x 96 pixels with
function menus |
| Backlight |
Two levels, normal and extra bright
for darker environments |
| Trigger lock |
Yes |
Switchable
Celsius and Fahrenheit |
Yes |
| Power |
2 'AA' LR6 Batteries (and USB powered
when used with a PC) |
| Battery life |
If used continuously: Laser and
backlight on, 12 hours;
Laser and backlight off, 100 hours. |
| Operating temperature |
0°C to 50°C
(32°F to 122°F) |
| Storage temperature |
-20°C to
60°C (-40°F
to 140°F) |
Bead thermocouple
Type-K range |
-40°C to
260°C (-40°F
to 500°F) |
Bead thermocouple
Type-K accuracy |
±1.1°C
(2.0°F) from 0°C
to 260°C (32°F
to 500°F), typically within 1.1°C
(2.0°F) from -40°C
to 0°C (-40°F
to 32°F) |