Description
The Fluke 700PTP is a handheld pressure pump designed to generate
either vacuum to -11.6 psi/-0.8 bar or pressure to 360 psi/25 bar.
The Fluke 700PTP has two pressure ports:
- 3/8-BSP (ISO228) female parallel thread
fitting for the reference gauge or pressure module
- 1/8-BSP (ISO228) female parallel thread
fitting for the unit under test.
Applications
The Fluke 700PTP features an integral pressure adjustment
vernier which varies the pressurized volume by 2.0 cc over
approximately eleven turns of the vernier knob. The pressure
variation achievable with the vernier will depend on the nominal
pressure and total pressurized volume, but with a minimum volume and
maximum pressure, the vernier provided 360 ¤20 psi adjustment range.
With a minimum volume and no pressure applied, the vernier can also
be used to provide a 0 to 70" H2O range. Larger volumes will provide
a smaller range of adjustment, but greater resolution. The length of
the stroke can be adjusted to limit the maximum output pressure.
Maximum output pressure is adjustable from 2.5 psi to 360 psi.
Included
 |
Adapter to
connect the reference gauge port on the pump body to the 1/4 NPT
male fitting on Fluke 700Pxx Pressure Modules (except 700P29,
P30, and P31). |
 |
One-meter
nylon hose to attach to unit under test port to the unit under
test. The hose terminates in a 1/4-BSP (ISO228) female parallel
thread fitting. |
 |
Adapter to
convert the 1/4-BSP female parallel thread fitting to a 1/4 NPT
female. |
 |
Replacement
seal kit. |
 |
Instruction sheet. |
Material: Nickel plated brass/anodized
aluminum.
Dimensions: 8.7" x 4.1" x 2.5" (220 mm x 105 mm x 63 mm)
Weight: 1.4 lb (0.65 kg)
Warranty: One year
|