| Fluke 717 Series |
Measure Pressure |
| |
Accuracy: |
| |
Functions: |
| |
|
|
|
(Internal sensor) |
|
0.05% full scale |
|
Zero, Min, Max, Hold, Damp |
|
Note: Gasses/Liquids
(non-corrosive) |
|
| Fluke 717 1G |
Range: |
| |
Resolution: |
| |
Over Pressure: |
|
|
-1 to 1 PSI, (-68.9 mbar to 68.9
mbar, -83 to 6.89 kPa) |
|
0.0001 PSI, 0.001 mbar, 0.0001
kPa |
|
5 x full scale |
|
| Fluke 717 30G |
Range: |
| |
Resolution: |
| |
Over Pressure: |
|
|
-12 to 30 PSI, (-850 mbar to 2
bar, -83 to 206.84 kPa) |
|
0.001 PSI, 0.1 mbar, 0.01 kPa |
|
2 x full scale |
|
| Fluke 717 100G |
Range: |
| |
Resolution: |
| |
Over Pressure: |
|
|
-12 to 100 PSI, (-850 mbar to 7
bar, -83 to 689.48 kPa) |
|
0.01 PSI, 0.1 mbar, 0.01 kPa |
|
2 x full scale |
|
| Fluke 717 300G |
Range: |
| |
Resolution: |
| |
Over Pressure: |
|
|
-12 to 300 PSI, (-850 mbar to 20
bar, -83 to 2068.42 kPa) |
|
0.01 PSI, 1 mbar, 0.1 kPa |
|
375 PSI, 25 bar |
|
| Fluke 717 500G |
Range: |
| |
Resolution: |
| |
Over Pressure: |
|
|
0 to 500 PSI, (0 mbar to 34.7
bar, 0 to 3447.38 kPa) |
|
0.01 PSI, 1 mbar, 0.1 kPa |
|
2 x full scale |
|
| Fluke 717 1000G |
Range: |
| |
Resolution: |
| |
Over Pressure: |
|
|
0 to 1000 PSI, (0 mbar to 69
bar, 0 to 6894.76 kPa) |
|
0.1 PSI, 1 mbar, 0.1 kPa |
|
2 x full scale |
|
| Fluke 717 1500G |
Range: |
| |
Resolution: |
| |
Over Pressure: |
|
|
0 to 1500 PSI, (0 mbar to 103.4
bar, 0 to 10342 kPa) |
|
0.1 PSI, 0.01 mbar, 1 kPa |
|
2 x full scale |
|
| Fluke 717 3000G |
Range: |
| |
Resolution: |
| |
Over Pressure: |
|
|
0 to 3000 PSI, (0 mbar to 207
bar, 0 to 20684.27 kPa) |
|
0.1 PSI, 0.01 mbar, 1 kPa |
|
2 x full scale |
|
| Fluke 717 5000G |
Range: |
| |
Resolution: |
| |
Over Pressure: |
|
|
0 to 5000 PSI, (0 mbar to 345
bar, 0 to 34474 kPa) |
|
0.1 PSI, 0.01 mbar, 1 kPa |
|
2 x full scale |
|
| Fluke 717 Series with Pressure |
| Modules, Measure Pressure |
|
Range: |
|
|
|
|
Resolution: |
|
|
Accuracy: |
|
|
Functions: |
|
| |
|
(via pressure module connector
using optional Pressure Module) |
|
29 pressure modules, 0 - 1 in H20
to 10,000 PSI, 2.5 mbar to 700 |
|
bar ranges. Over pressure per
pressure module specifications. |
|
Per pressure module specs |
|
Per pressure module specs |
|
Zero, Min, Max, Hold, Damp |
|
|
| |
| |
 |
| |
| Current |
Range: |
| |
Resolution: |
| |
Accuracy: |
|
|
0 to 24 mA (-25% to 125%) |
|
0.001 mA |
|
0.015% + 2 counts |
|
| Loop Power |
Range: |
| |
Accuracy: |
| |
|
| |
|
|
|
24v dc |
|
±10% |
|
Note: Drive; 20mA into 1000Ω
for battery > 6.8V; |
|
700W for battery 5.8 to 6.8V |
|
| Display |
Clear 5 digit LCD |
|
|
|
|
|
 |
|
|
|
Operating temperature |
-10șC to
55șC |
|
Storage temperature |
-40șC to
60șC |
|
Relative humidity (% RH operating without condensation): |
95% (10șC
to 30șC); 75% (30șC to 40șC);
45% (40șC to 50șC); 35% (50șC to 55șC) |
|
Operating altitude |
3,000m max |
|
Vibration |
Random,
2g, 5-500 Hz |
|
Shock |
1m drop
test |
|
Safety |
CSA C22.2
No. 1010.1:1992 |
|
Power |
Single 9V
battery ANSI/NEDA 1604A
or IEC 6LR619V alkaline |
|
Size
(with holster) |
201 mm L x
98mm W x 52 mm D
(7.93 in X 3.86 in x 2.06 in) |
|
Weight
(with holster) |
Approx.
600g (21 oz) |
|
Size
(without holster) |
187 mm L x
87mm W x 32 mm D
(7.35 in X 3.41 in x 1.25 in) |
|
Weight
(without holster) |
Approx.
330g (12 oz) |
|
Warranty |