Leda Electronics Logo. Western  Australia's largest distributor of Test Instruments Leda Electronics Showing The Way

 
 
Fluke Test Instruments Logo

Fluke Process Calibration Instruments

 
Model 726 Precision Multifunction Process Calibrator - Specifications  
 
 
Model 726 Measurement Accuracy
Voltage DC 30.000 V
  20.000 V
  90.00 mV
  -10mV to 75.00mV
0.01% + 2mV (upper display)
0.01% + 2mV (lower display)
0.01% + 2mV to 10mV
0.025% + 1 count (via TC connection)
Current DC

24.000 mA

0.01% + 2mA
Resistance 0.0 to 4000.0 Ω
±0.015%
Frequency 2.0 to 1000.0 CPM
  1.0 to 1000.0 Hz
  1.00 to 15.00 kHz
  Sensitivity
0.05% + 0.1 cpm
0.05% + 0.1 Hz
0.05% + 0.1 kHz
1V peak-to-peak-minimum
Pressure

Accuracy

from 0.025% of range using any of 29 pressure modules.*
Note: * for detailed specifications refer to pressure modules in Options and Accessories.
           Modules available for differential, gage, vacuum, absolute, dual and high pressure.
 
 
Model 726 Source Accuracy
Voltage DC 100.00 mV
  20.000 V
  -10mV to 75.00mV
0.01% + 10mV
0.01% + 2mV
0.025% + 1 count (via TC connection)
Current DC

24.000 mA (Source)

24.000 mA (Simulate)

0.01% + 2mA
0.01% + 2mA
Resistance
 
 
   
   
   
   
   
Range
 
Accuracy
 
Excitation
Current
5Ω to 400Ω 0.015% + 0.1Ω 0.1mA to 0.5mA
5Ω to 400Ω 0.015% + 0.05Ω 0.5mA to 3mA
400Ω to 1500Ω 0.015% + 0.5Ω 0.5mA to 0.8mA
1500Ω to 4000Ω 0.015% + 0.5Ω 0.5mA to 0.4mA
Percent of reading + floor
Frequency 2.0 to 1000.0 CPM
  1.0 to 1000.0 Hz
  1.0 to 10.0 kHz
  10.0 to 15.0 kHz
  Waveform
0.05%
0.05%
0.25%
0.5%
5V p-p square-wave, -0.1V offset
 
 
RTD's and Thermocouples
Measure and Source Accuracy
  Cu10
  NI-120
  PT-100 (385)
  PT-100 (392)
  PT-100 (391-JIS)
  PT-200 (385)
  PT-500 (385)
  PT-1000 (385)
 

Resolution:

 

J:

 

K:

 

T:

 

E:

 

R:

 

S:

 

B:

 

L:

 

U:

 

N:

 

XK:

 

BP:

   

1.8°C

-10°C to 250°C

0.15°C

-80°C to 260°C

0.25°C

-200°C to 800°C

0.15°C

-200°C to 630°C

0.25°C

-200°C to 630°C

0.75°C

-200°C to 630°C

0.35°C

-200°C to 630°C

0.15°C

-200°C to 630°C

0.01°C

 

0.4°C

 

0.5°C

 

0.4°C

 

0.4°C

 

1.4°C

 

1.4°C

 

1.4°C

 

0.4C

 

0.4C

 

0.6°C

 

0.4°C

 

1.1°C

 
 
 
Thermocouple Resolution
J, K, T, E, L, N, U 0.1ºC, 0.1ºF
B, R, S 1ºC, 1ºF
 
 
Specifications
Ramp Functions
  Source Functions
  Ramps
 
Voltage, current, resistance, frequency, temperature
Slow ramp, Fast ramp, 25% step-ramp
Step Functions
  Source Functions
  Steps
 
Voltage, current, resistance, frequency, temperature
25% of range, 100% of range
Loop Power Function
  Voltage
  Accuracy
  Maximum Current
 
24V
10%
22mA (short circuit protected)
 
 
Mechanical and General Specifications
Storage Temperature -20ºC to 71ºC
Operating Temperature -10ºC to 55ºC
Safety Approvals EN 61010-1: 1993, ANSI/ISA S82.01-1994
CSA C22.2 No. 1010.1:1992
Operating Altitude 3000 meters
Size 200mm x 96mm x 47mm (7.9" x 3.8" x 1.9")
Weight 650g (23oz)
Battery Four AA alkaline batteries.
Battery Replacement Separate battery compartment, accessible without breaking calibration seal.
Side Port Connections Pressure module connector, also used for remote real-time programming.

 

Home | Amprobe | CIE | EuroSMC | Fluke | Kyoritsu-Kew |Megger | Reed | Yokogawa | Site Map | Search

Leda Electronics Pty. Ltd.
A.C.N. 078 681 562           A.B.N. 83 078 681 562
P.O. Box 1733 Malaga, LPO WA 6944
                                                      Unit 5, 31 Century Road, Malaga, Western Australia 6090                                                            
Phone: 61 8 9248 6688    Fax: 61 8 9248 6227

information@ledaelectronics.com.au

          Copyright © 2000 - 2011                                                                Last Modified: Thursday, 15 September 2011