Leda Electronics Logo. Western  Australia's largest distributor of Test Instruments Leda Electronics Showing The Way

 
 
Fluke Test Instruments Logo

Fluke Process Calibration Instruments

 
Model 743 Documenting Process Calibrator - Specifications  
 
 
Model 743 Source Accuracy
 
Range (full scale) Accuracy
(% of reading + % of full scale)
110.000 mV 0.01% + 0.005%
1.10000V 0.01% + 0.005%
15.0000V 0.01% + 0.005%
Source 22.000 mA 0.01% + 0.015%
Simulate 22.000 mA 0.02% + 0.03%
11.000W 0.01% + 20 mW
110.00W 0.01% + 40 mW
1.1000 kW 0.02% + 0.5W
11.000 kW 0.03% + 5W
0.00 to 10.99 Hz 0.01 Hz
11.00 to 109.99 Hz 0.1 Hz
110.0 to 1099.9 Hz 0.1 Hz
1.100 to 21.999 kHz 2 Hz
22.000 to 50.000 kHz 5 Hz
 
 
Model 743 Measurement Accuracy
 
Range (Full scale) Accuracy
(% of reading + % of full scale)
110.000 mV DC 0.025% + 0.015%
1.10000V DC 0.025% + 0.005%
11.0000V DC 0.025% + 0.005%
110.000V DC 0.05% + 0.005%
300.00V DC 0.05% + 0.005%
V AC, 20 to 40 Hz 2% + 10 counts
V AC, 40 to 500 Hz 0.5% + 5
V AC, 500 to 1 kHz 2% + 10
V AC, 1 kHz to 5 kHz 10% + 20
30.000 mA DC 0.01% + 0.015%
110.00 mA DC 0.01% + 0.015%
11.000W 0.05% + 50 mW
110.00W 0.05% + 50 mW
1.1000 kW 0.05% + 0.5W
11.000 kW 0.1% + 10W
1.00 to 109.99 Hz 0.05 Hz
110.0 to 1099.9 Hz 0.5 Hz
1.100 to 10.999 kHz 5 Hz
11.00 to 50.00 kHz 50 Hz
 
 
Model 743 RTD and Thermocouple Specifications
 
Device Specification
10W Cu (427) Measure Accuracy: 3°C
Source Accuracy: 1°C
100W Pt (3916) Measure Accuracy: 0.3°C
Source Accuracy: 0.1°C
100W Pt (3926) Measure Accuracy: 0.3°C
Source Accuracy: 0.1°C
100W Pt (385) Measure Accuracy: 0.3°C
Source Accuracy: 0.1°C
200W Pt (385) Measure Accuracy: 0.3°C
Source Accuracy: 0.1°C
500W Pt (385) Measure Accuracy: 0.3°C
Source Accuracy: 0.1°C
1000W Pt (385) Measure Accuracy: 0.3°C
Source Accuracy: 0.1°C
120W Ni (672) Measure Accuracy: 0.3°C
Source Accuracy: 0.1°C
E Measure Accuracy: 0.3°C
Source Accuracy: 0.2°C
N Measure Accuracy: 0.5°C
Source Accuracy: 0.3°C
J Measure Accuracy: 0.3°C
Source Accuracy: 0.2°C
L Measure Accuracy: 0.3°C
Source Accuracy: 0.2°C
K Measure Accuracy: 0.3°C
Source Accuracy: 0.3°C
T Measure Accuracy: 0.3°C
Source Accuracy: 0.3°C
U Measure Accuracy: 0.3°C
Source Accuracy: 0.3°C
B Measure Accuracy: 0.9°C
Source Accuracy: 0.8°C
R Measure Accuracy: 1.0°C
Source Accuracy: 0.9°C
S Measure Accuracy: 0.9°C
Source Accuracy: 0.9°C
C Measure Accuracy: 0.6°C
Source Accuracy: 0.6°C

Note*Resolution: 0.1°C, except 1°C for 10W Cu. Best case, mid-range accuracies. Sensor inaccuracies not included. For 2-wire and 3-wire RTD measurements, add 0.4°C. Thermocouple accuracies with external cold junction. For internal junction add 0.2°C.

 
 
Model 743 Mechanical and General Specifications
 
Dimensions 130 x 236 x 61 mm
Weight 1.4 kg (3 lb 1 oz)
Internal Battery Pack NiCd, 7.2V, 1700 mAh
Battery Life Typical usage > eight hours
Battery Replacement Via snap-shut door without opening calibrator; no tools required
Side Port Connections
  • Pressure module connector
  • RS-232 connector to interface to your PC
  • Connection for optional battery eliminator
Safety Complies with CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993
Data Storage Capacity 1 week of calibration results
90 Day Specifications

The standard specification interval for the 743B is 1 or 2 years. Typical 90 day measurement and source accuracy can be estimated by dividing the one year "% of reading" or "% of output" specifications by 2. Floor specifications, expressed as "% of f.s.", "counts" or "ohms", remain constant.

 

Home | Amprobe | CIE | EuroSMC | Fluke | Kyoritsu-Kew |Megger | Reed | Yokogawa | Site Map | Search

Leda Electronics Pty. Ltd.
A.C.N. 078 681 562           A.B.N. 83 078 681 562
P.O. Box 1733 Malaga, LPO WA 6944
                                                      Unit 5, 31 Century Road, Malaga, Western Australia 6090                                                            
Phone: 61 8 9248 6688    Fax: 61 8 9248 6227

information@ledaelectronics.com.au

          Copyright © 2000 - 2011                                                                Last Modified: Thursday, 15 September 2011