787 ProcessMeter™ - Specifications  
 
 
Model 787 ProcessMeter Specifications
DC Current Output (Internal battery operation) Range and Resolution
0.000 to 20.000 mA or 4.000 to 20.000 mA (selectable at power-up). Over-range to 24.000 mA
Drive Capability

12 V compliance, or 500 ohms @20 mA
Accuracy (% of span)

0.05 %
DC Current Simulate (External 24 volt loop supply) Range and Resolution
0.000 to 20.000 mA or 4.000 to 20.000 mA (selectable at power-up). Over-range to 24.000 mA
Drive Capability

1000 ohms @20 mA
Accuracy (% of span)

0.05 %
Current Adjustment Modes Range and Resolution
Manual: Coarse Fine: 25% step
Automatic: Slow Ramp, Fast Ramp, 25% step
 
 
Model 787 ProcessMeter Measurement Accuracy
VDC Range and Resolution:
400.0 mV, 4.000V, 40.00V, 400.0V, 1000V
Best Accuracy (% of reading + LSD):

0.1% + 1
VAC (true RMS)
to 1 kHz
*
Range and Resolution:
400.0 mV, 4.000V, 40.00V, 400.0V, 1000V
Best Accuracy (% of reading + LSD):

0.7% + 2
mA DC Range and Resolution:
30.000 mA
Best Accuracy (% of reading + LSD:

0.05% + 2
A DC Range and Resolution:
1.000 A (0.440 A continuous)
Best Accuracy (% of reading + LSD):

0.2% + 2
A AC Range and Resolution:
1.000 A (0.440 A continuous)
Best Accuracy (% of reading + LSD):

1% + 2
Resistance Range and Resolution:
400.0 Ohms, 4.000 kOhms, 40.00 kOhms, 400.0 kOhms, 4.000 MOhms, 40.00MOhms
Best Accuracy (% of reading + LSD):

0.2% + 1
Frequency (0.5 Hz to 20 kHz) Range and Resolution:
199.99 Hz, 1999.9 Hz, 19.999 kHz, 199.99 kHz
Best Accuracy (% of reading + LSD):

0.005% + 1
Diode Test Range and Resolution:
2.400 V (measures diode voltage drop)
Best Accuracy (% of reading + LSD):

2% + 1
Continuity Range and Resolution
Beeps for resistance < approx. 100 ohms
 
 
Model 787 ProcessMeter Mechanical and General Specifications
Maximum voltage applied between any jack and Earth Ground 1000 V rms
Storage Temperature -40 ºC to 60 ºC
Operating Temperature -20 ºC to 55 ºC
Temperature Coefficient 0.05 x (specified accuracy) per ºC (for temperatures < 18 ºC or > 28 ºC)
Relative Humidity 95% up to 30 ºC; 75% up to 40 ºC; 45% up to 50 ºC; 35% up to 55 ºC
Vibration Random, 2g, 5-500 Hz
Shock 1 meter drop test
Safety Designed in accordance with IEC 1010-1, ANSI/ISA S82.01-1994 and CAN/CSA C22.2 No. 1010.1-92 Overvoltage Category III
Size 32 mm H x 87 mm W x 187 mm L (1.25 in H x 3.41 in W x 7.35 in L)
Size, with holster and Flex-Stand 52 mm H x 98 mm W x 201 mm L (2.06 in H x 3.86 in W x 7.93 in L)
Weight 369 g (13 oz); with holster and Flex-Stand 638 g (22.5 oz)
Battery Single 9V Alkaline battery (ANSI/NEDA 1604A or IEC 6LR61)
Battery Life 50 hours typical (measurement), 12 hours typical (sourcing 12 mA)
Warranty 3 years

 

Home | Amprobe | CIE | EuroSMC | Fluke | Kyoritsu-Kew |Megger | Reed | Yokogawa | Site Map | Search

Leda Electronics Pty. Ltd.
A.C.N. 078 681 562           A.B.N. 83 078 681 562
P.O. Box 1733 Malaga, LPO WA 6944
                                                      Unit 5, 31 Century Road, Malaga, Western Australia 6090                                                            
Phone: 61 8 9248 6688    Fax: 61 8 9248 6227

information@ledaelectronics.com.au

          Copyright © 2000 - 2011                                                                Last Modified: Thursday, 15 September 2011