Voltage DC
|
Maximum voltage: Accuracy: Maximum resolution: |
1000V
±(0.05% +1) 10µV |
Voltage AC
|
Maximum voltage: Accuracy: AC bandwidth: Maximum resolution: |
1000V
±(0.7% +2) True-RMS 20kHz* 0.1mV |
Current DC
|
Maximum amps: Accuracy: Maximum resolution: |
10A (20A for 30 seconds max)
±(0.02% +2) 0.01µA |
Current AC
|
Maximum amps: Accuracy: Maximum resolution: |
10A (20A for 30 seconds max)
±(1.0% +2) True-RMS 0.1µA |
Resistance
|
Maximum resistance: Accuracy: Maximum resolution: |
50MΩ ±(0.2% +1) 0.1Ω |
Capacitance
|
Maximum capacitance: Accuracy: Maximum resolution: |
9.999µF
±(1% +2) 0.01nF |
Frequency
|
Maximum frequency: Accuracy: Maximum resolution: |
200kHz
±(0.005% +1) 0.01Hz |
Duty Cycle
|
Maximum duty cycle: Accuracy: Maximum resolution: |
99.90%
±(0.2% per kHz +0.1%) 0.10% |
Temperature Measurement |
Range: |
-200.0°C
to 1090°C (-328°F to 1994°F) |
80BK Temperature Probe
|
Range:
Accuracy: |
-40°C to
260°C (-40°F to 500°F) 2.2C or 2% whichever is greater |
Conductance
|
Maximum conductance Accuracy: Maximum resolution: |
60.00nS
±(1% +10) 0.01nS |
Diode
|
Range: Accuracy: Resolution: |
3V ±(2% +1) 1mV |
| Duty Cycle Range |
Accuracy: |
Within ±(0.2%
per kHz +0.1%) |
| Measurements Restriction in
Ex-hazardous area |
Voltage: Current: |
≤65V
≤5A |
Environmental Specifications
|
Operating temperature: Storage temperature: Humidity
(without condensation):
Operating altitude: |
-20°C to
+55°C -40°C to +60°C 0% to 90% (0°C to 35°C) 0% to 70% (35°C to 55°C) 2000m |
Safety Specifications
|
Overvoltage category:
Agency approvals: ATEX specification: Agency approvals: |
EN 61010-1 to 1000V CAT III, 600V CAT IV. CSA, TUV EX II 2 G EEx ia IIC T4 ZELM 05 ATEX 027 |
Mechanical and General
Specifications
|
Size including holster: Weight: Weight with holster: Display (digital):
Display (analog): Battery life (Alkaline):
Shock:
Vibration:
Warranty: |
201mm x 98mm x 52mm 1.09 lbs 1.46 lbs 6000 counts updates 4/sec 19,999 counts in high-resolution mode. 32 segments, updates 40/sec 400 hours typical, without backlight 1 meter drop
per IEC 61010-1:2001 Per MIL-PRF-28800 for a class 2 instrument 1 year |