Leda Electronics Logo. Western  Australia's largest distributor of Test Instruments Leda Electronics Showing The Way

 
Yokogawa > Other Test and Measurement Instruments > MT10 Series >   Ordering Codes/Accessories
Yokogawa Test and Measurement Logo

Other Test and Measurement Instruments ● Laboratory Pen Recorders ● Manometers ● Digital Bench Multimeters ● Time Interval Analyzers ● Universal Counters ● Precision Digital Laboratory Thermometers

MT10 Series Mini Manometer

Overview
Compared to the conventional metal sensor, it has high reproducibility and long term stability of characteristics, thus ensuring highly stable measurement. The MT10 is small and has a mass of about 700g. This makes it suitable for use at process industry sites, for diagnosing factory equipment and for various R & D applications.
 

MT10 Series Mini Manometer

Features
  • High reliability (with silicon resonant sensor).
  • Available in three models:
    265302 (130kPa)
    265303 (700 kPa)
    265304 (3000kPa)
  • Resolution 0.01kPa, 0.1kPa, 1kPa.
  • Data hold function.
  • Zero adjustment function.
  • RS-232C interface.
  • Supplied with: Carrying case, Battery installed, Monkey spanner, Single-ended wrench, Duct tape, and user's manual.

MT10 Series Datasheet (PDF)
(documents are in PDF format and will open in a new window)

Silicon-Resonant Sensor

Silicon-Resonant Sensor
 

The pressure sensor employs a silicon-resonant sensor which was developed using Yokogawa's newest silicon micro-forming technology.

The detector, which uses a silicon monocrystal, has superior reproducibility and undergoes less secular change than conventional metal sensors, thus permitting highly stable measurement.

MT10 Series Specifications

Measurement

Model 265302

Model 265303

Model 265304

Kind of Pressure

Gauge pressure

Gauge pressure

Gauge pressure

Range
(Accuracy Guarantee Range)

0 to 130kPa

0 to 700kPa

0 to 3000kPa

0 to 1.3256kgf/cm2

0 to 7.138kgf/cm2

0 to 30.59kgf/cm2

0 to 13256mmH2O

―――

―――

0 to 975.1mmHg

0 to 5250mmHg

―――

0 to 521.9 inH2O

0 to 2810 inH2O

0 to 12044 inH2O

0 to 38.39 inHg

0 to 206.7 inHg

0 to 885.9 inHg

0 to 18.85 psi

0 to 101.5 psi

0 to 435.1 psi

0 to 1300 mbar

0 to 7000 mbar

0 to 30.00 bar

Measurement Indication Range

-2.5 to 110% of FS

-2.5 to 110% of FS

-2.5 to 110% of FS

Accuracy (Horizontally mounted, 23°C ±3°C, zero point adjusted)

±(0.04% of rdg
+0.03% of FS)

±0.1% of FS

±0.1% of FS

Resolution

0.01kPa

0.1kPa

1kPa

Maximum Allowable Input

500kPa

1000kPa

4500kPa

Internal Volume

approx. 2cm3

approx. 2cm3

approx. 2cm3

Effect of Temperature

±0.02% of FE/10°C max. (Zero point)
±0.02% of FE/10°C max. (Span)

Effect of Mounting Direction

±0.1% of FS

±0.02% of FS

±0.01% of FS

Gas Leakage

10-5cm3/s max

Measurement Fluid

Gas only (non-flammable, non-explosive, non-toxic, non-corrosive)

Measurement Fluid Temperature

5°C to 50°C

5°C to 50°C

5°C to 50°C

Pressure Sensor

Silicon resonant sensor

Pressure Sensing Element

Diaphragm

Pressure Display Unit

kPa, kg/cm2, mmH2O, mmHg, inH2O, inHg, psi, mbar (bar)

Calibration Interval

1 year

Input Connection

Rc 1/4 or NPT 1/4

Material of Measurement Section

SUS316 stainless steel, Hastelloy C726

 
 

General Specifications

Display

LCD display

Display Update Interval

approx. 0.5 sec.

Response Time

approx. 4 seconds (until the accuracy falls within the accuracy range)

Warm-up Time

1 minute maximum

Other Functions

Comes with data hold function that holds indicated pressure value, and zero adjustment function.

Operating Temperature

5°C to 40°C; 20% to 80% RH or less (no condensation)

Power Supply

4 x 1.5 Volt Type 'AA' Alkaline batteries

Battery Life

approx. 100 hours (continuous operation using alkaline batteries)

AC Adapter

100 to 120/220 to 240 Volts AC 50/60Hz (output 15V, 1.33A) (optional)

Communication Interface

RS-232C Half-duplex transmission brain protocol. Transmission rate: 1200 bits/s

Insulation Resistance

500V DC/20MΩ or higher (between AC line and case) (when dedicated power kit is used).

Withstand Voltage

1500VAC/minute (between AC line and case) (when dedicated power kit is used).

Dimensions

approx. 72mm (W) x 174mm (H) x 60mm (D) (excluding input connection section)

Weight

approx. 700g (including battery

Home | Amprobe | CIE | EuroSMC | Fluke | Kyoritsu-Kew |Megger | Reed | Yokogawa | Site Map | Search

Leda Electronics Pty. Ltd.
A.C.N. 078 681 562           A.B.N. 83 078 681 562
P.O. Box 1733 Malaga, LPO WA 6944
                                                      Unit 5, 31 Century Road, Malaga, Western Australia 6090                                                            
Phone: 61 8 9248 6688    Fax: 61 8 9248 6227