MVAdvanced, MV1000 and MV2000 Series
Paperless
Recorders Functions
|
 |
| Input Components |
|
Number of
Inputs |
Series MV1000 |
4, 6,8, 12 or
24 channels |
|
Series MV2000 |
8, 10, 20, 30,
40 or 48 channels |
|
Measurement
Intervals
(A/D
integration time is fixed at 1.67ms in High-Speed mode) |
Model MV1004 |
125ms, 250ms
or 25ms in High-Speed mode |
|
Model MV1008 |
|
Model MV2008 |
|
Model MV1006 |
1s, 5s or
125ms in High-Speed mode |
|
Model MV1012 |
|
Model MV1024 |
|
Model MV2010 |
|
Model MV2020 |
|
Model MV2030 |
|
Model MV2040 |
|
Model MV2048 |
|
Points to
consider when using
High-Speed Mode
|
When using
High-Speed mode (an A/D integration time of 1.67ms) with the
MVAdvanced, power supply noise and other factors may cause the
measured values to fluctuate.
If this is the
case, then measure using Normal mode (A/D integration time of
16.7ms, 20ms or 100ms). |
|
| |
| |
| Input Types |
|
DC Voltage |
2omV, 60mV,
200mV, 2V, 6V, 20V, 50V and 1 to 5V |
|
Thermocouple
Types |
R, S, B, K, E,
L, T, N, W, L, U, WRe |
|
RTD Types |
Pt100, JPt100 |
|
DI |
at the contact
input or the TTL level |
|
DC Current |
with external
shunt resistor |
|
| |
|
Measurement: Range, Accuracy, and
Display Resolution |
|
Input
|
Range and
Type |
Measurement
Accuracy
(when the integration time is 16.7ms or more) |
Display
Resolution
|
|
DC Voltage |
1-5V |
±(0.05% of reading + 3
digits) |
1mV |
|
Thermocouple* |
'K' |
±(0.15% of reading + 0.7°C) |
0.1°C |
|
RTD's |
'Pt100' |
±(0.15% of reading + 0.3°C) |
0.1°C |
|
| Note: * Does not include the accuracy of
reference junction compensation |
| |
| Display |
|
Display
Device |
Series MV1000 |
5.5-inch TFT
color LCD (320 x 240 dots) |
|
Series MV2000 |
10.4-inch TFT
color LCD (640 x 480 dots) |
|
Note: the LCD
may contain some pixels that are always lighted or that never
light, and variations in brightness may occur due to the
characteristics of liquid crystals. Please note that these are
not defects. |
|
Display
Groups |
Number of
display groups: |
|
Series MV1000 |
10 groups |
|
Series MV2000 |
36 groups |
|
Maximum
number of channels assigned per group: |
|
Series MV1000 |
6 channels |
|
Series MV2000 |
10 channels |
|
Display
Colors |
Trend/bar |
24 available
colors |
|
Background |
white or black |
|
Trend
Display |
Display Types |
Vertical,
horizontal,, horizontal wide, separated horizontal |
|
Bar Graph
Display |
Direction |
Vertical or
horizontal |
|
Digital
Display |
Update rate |
1 second |
|
Overview
Display |
Channel number |
Displays a
list of all measurement and MATH channels along with their alarm
states. |
|
Information
display |
Alarm Summary,
Message Summery, Memory Summery, Report, Relay and Modbus
Condition. |
|
Log Display |
Display
content |
Log-in, Error,
Transmission, FTP, Web, E-mail, SNTP, DHCP, Modbus |
|
Tag Display |
Maximum
displayable characters: 16 alphanumeric |
|
Message
Display |
Maximum
displayable characters: 32 alphanumeric |
|
Historical
display function: Allows for the display of data stored to
internal or external memory. |
|
LCD Screen
Saver Function |
You can DIM or
turn OFF the LCD backlight if there are no keystrokes for a set
time (set times: 1, 2, 5, 10, 30 min or 1 hour) |
|
Display
Screen Registration |
You can assign
a name to a display screen and register it. Maximum registration
screens: 8 |
|
| |
| |
| Memory Functions |
|
External
Media |
Media |
Compact Flash
(CF) card |
|
Internal
Memory |
Media |
Flash memory |
|
Memory size |
80MB or 200MB
(selected when purchasing) |
|
| |
|
Save
Interval |
125ms |
1s |
5s |
10s |
60s |
120s |
|
Sample Time
(with 200MB of memory) |
Approx.
217 hours |
Approx.
72 days |
Approx.
362 days |
Approx.
723 days |
Approx.
12.1 years |
Approx.
24 years |
|
Sample Time
(with 80MB of memory) |
Approx.
86 hours |
Approx.
28 days |
Approx.
144 days |
Approx.
289 days |
Approx.
4.8 years |
Approx.
9.6 years |
|
| Examples of
internal memory sample times (with the Model MV1012 recording only event
data files for 12 measuring channels and no calculation channels). |
| |
|
Maximum
Savable Data Files |
400 (savable
display data files and event data files combined) |
|
Manual Save |
Saves data files
to the internal memory manually. You can save all data or only
selected data to a Destination drive (CF card or USB memory). |
|
Auto Save |
Save displayed
data to the CF card at set interval.
Save event data to the CF card at a set interval (in Free Trigger
mode) Save when finished sampling (when setting the trigger). |
|
Data Formats |
When saving to
external media, both event data and display data can be saved in
either binary or text format (data is always stored to internal
memory in binary format). |
|
Manual Sample
Data |
You can choose
when to save measurement/ MATH channel data files to the internal
memory or to a CF card. |
|
Report Data |
Saves
calculation results as data to a CF card at set times. (only
available if the unit is equipped with the calculation option). |
|
Types |
Hourly, daily,
hourly + daily, weekly, daily + monthly |
|
Trigger
Function |
Data can be
saved using Free mode or Trigger mode. When using Trigger mode,
the user must set the data length, pre-trigger, and trigger
source. |
|
Snapshot
Function |
Saves the
displayed screen image data to a CF card. |
|
Data File
Loading |
Data files saved
to a CF card or to a USB memory (if equipped with the USB option)
can be loaded and displayed.
Loading and
saving setup data: Settings data can be saved and loaded in Binary
format. |
|
USB Interface |
USB
specification 1.1 host |
|
Ports |
Two (front and
back) |
|
Connection
Devices |
104 keyboards
(US) compliant with USB HID Class Version 1.1
109 keyboards (Japanese)
External media: USB memory (not all types of USB memory are
guaranteed to work). |
|
| |
| |
| |
| |
|