Leda Electronics Logo. Western  Australia's largest distributor of Test Instruments Leda Electronics Showing The Way

 
Model Series MX100 > Benefits > Software > MX100 Features > 1 > 2 > 3 > 4 > 5 >
Yokogawa Test and Measurement Logo

PC Based Data Acquisition

Model Type MX100
PC-Based Data Acquisition Unit Type MX100 - Features Page 4 of 5

Hardware Configuration

The MX is designed to enable desired measurement environments by combining three elements: the main module, input/output modules, and a base plate.
The assembled unit can be utilized on the desktop as-is (the modules have legs). Please note that DIN rails are used for rack-mounting.

Module

 
Main Module (MX100)

The main module is the engine that controls data acquisition. It is equipped with a power supply, an Ethernet port, and a Compact Flash card slot. One main module can accommodate up to six input/output modules. The user can choose up to six modules of any type.
 

  Data Backup Time by CF Card Size
(Note: Minutes/hours/days are approximate.)

 
main module
Ethernet port:
For communication with the PC. Automatic recognition of 100Base-TX or 10Base-T.
CF (Compact Flash) Card Slot:
CF
Number of
saved channels
Measured
interval
32M 128M 512M
10 channels 10ms 2 hours 9 hours 36 hours
100ms 21 hours 3.5 days 15 days
500ms 4.5 days 18.5 days 75.5 days
2sec 18 days 75.5 days 303 days
24 channels 10ms 54 minutes 3.5 hours 15 hours
100ms 9 hours 37 hours 6 days
500ms 45 hours 7.5 days 31.5 days
2sec 7.5 days 31.5 days 126 days
60 channels 10ms 18 minutes 75 minutes 5 hours
100ms 3.5 hours 15 hours 2.5 days
500ms 18 hours 3 days 12.5 days
2sec 3 days 12.5 days 50 days

With the CF Card, measured data is saved to the card if communication with the PC is interrupted. Since the card provides data backup, you do not have to worry about the possible loss of data even if the MX is configured for PC-based measurement.
 

Input/Output Modules (MX110, MX115, MX125)
Note:
All I/O modules are fitted with a swing-up terminal cover.

 
Four-Channel High-Speed Universal Input Module:
MX110-UNV-H04
MX110-UNV-H04

MX110-UNV-H04
Shortest measurement interval: 10 ms; Maximum number of input channels: 4 channels; Mixed input of DC voltage, thermocouple, RTD, and digital input: possible. Please note that the shortest measurement interval for temperature is 50 ms for this module.


 


 
Ten-Channel Medium-Speed Universal Input Module:
MX110-UNV-M10
MX110-UNV-M10

MX110-UNV-M10
Shortest measurement interval: 100 ms; Maximum number of input channels: 10 channels; Mixed input of DC voltage, thermocouple, RTD, and digital input: possible.




 


 
Ten-Channel High-Speed Digital Input Module
MX115-D05-H10
MX115-D05-H10

MX115-D05-H10
If a single unit consists of this module only, measurement of 60 channels/10 ms is possible. Most suitable for measurement of digital input signals at many points inexpensively. Shortest measurement interval: 10 ms; Maximum number of input channels: 10 channels; Digital input dedicated module.

 


 
Ten-Channel Medium-Speed Digital Output Module:
MX125-MKC-M10
MX125-MKC-M10

MX125-MKC-M10
For use of alarm outputs, etc. Shortest output interval: 100 ms; Maximum number of output channels: 10 channels; Digital output dedicated module.





 


 

Home | Amprobe | CIE | EuroSMC | Fluke | Kyoritsu-Kew |Megger | Reed | Yokogawa | Site Map | Search

Leda Electronics Pty. Ltd.
A.C.N. 078 681 562           A.B.N. 83 078 681 562
P.O. Box 1733 Malaga, LPO WA 6944
                                                      Unit 5, 31 Century Road, Malaga, Western Australia 6090                                                            
Phone: 61 8 9248 6688    Fax: 61 8 9248 6227

information@ledaelectronics.com.au

          Copyright © 2000 - 2011                                                                Last Modified: Tuesday, 15 November 2011