| Hardware Configuration |
|
The MX is designed to
enable desired measurement environments by combining three elements: the
main module, input/output modules, and a base plate. The assembled unit can be utilized on the desktop as-is (the modules have
legs). Please note that DIN rails are used for rack-mounting. |
 |
| |
| Main Module (MX100) |
|
The main module is the engine that controls data
acquisition. It is equipped with a power supply, an Ethernet port, and a
Compact Flash card slot. One main module can accommodate up to six
input/output modules. The user can choose up to six modules of any type. |
| |
Data Backup Time by CF Card Size
(Note: Minutes/hours/days are approximate.)
|

Ethernet port:
For communication with the PC. Automatic recognition of 100Base-TX or
10Base-T.
CF (Compact Flash) Card Slot:
 |
Number of
saved channels |
Measured
interval |
32M |
128M |
512M |
|
10 channels |
10ms |
2 hours |
9 hours |
36
hours |
|
100ms |
21
hours |
3.5
days |
15 days |
|
500ms |
4.5
days |
18.5
days |
75.5
days |
|
2sec |
18 days |
75.5
days |
303
days |
|
24
channels |
10ms |
54
minutes |
3.5
hours |
15
hours |
|
100ms |
9 hours |
37
hours |
6 days |
|
500ms |
45
hours |
7.5
days |
31.5
days |
|
2sec |
7.5
days |
31.5
days |
126
days |
|
60
channels |
10ms |
18
minutes |
75
minutes |
5 hours |
|
100ms |
3.5
hours |
15
hours |
2.5
days |
|
500ms |
18
hours |
3 days |
12.5
days |
|
2sec |
3 days |
12.5
days |
50 days |
|
|
With the CF Card, measured data is saved to the card if
communication with the PC is interrupted. Since the card provides data backup,
you do not have to worry about the possible loss of data even if the MX is
configured for PC-based measurement.
|
Input/Output Modules (MX110, MX115, MX125)
Note:
All I/O modules are fitted with a swing-up terminal cover.
|
Four-Channel High-Speed Universal Input
Module:
MX110-UNV-H04
 |
MX110-UNV-H04
Shortest measurement interval: 10 ms; Maximum number of input channels: 4
channels; Mixed input of DC voltage, thermocouple, RTD, and digital input:
possible. Please note that the shortest measurement interval for
temperature is 50 ms for this module.
|
|
|
Ten-Channel Medium-Speed Universal Input
Module:
MX110-UNV-M10
 |
MX110-UNV-M10
Shortest measurement interval: 100 ms; Maximum number of input
channels: 10 channels; Mixed input of DC voltage, thermocouple, RTD, and
digital input: possible.
|
|
|
Ten-Channel High-Speed Digital Input Module
MX115-D05-H10
 |
MX115-D05-H10
If a single unit consists of this module only, measurement of 60
channels/10 ms is possible. Most suitable for measurement of digital input
signals at many points inexpensively. Shortest measurement interval: 10
ms; Maximum number of input channels: 10 channels; Digital input dedicated
module.
|
|
|
Ten-Channel Medium-Speed Digital Output
Module:
MX125-MKC-M10
 |
MX125-MKC-M10
For use of alarm outputs, etc. Shortest output interval: 100 ms;
Maximum number of output channels: 10 channels; Digital output dedicated
module.
|
|