| WT3000 Controls
are easy to use, easy to view The
WT3000 was designed with user-friendly functions and controls in
response to user requests for a simpler range setting operation
and more user-friendly parameter setting display process.
- Range settings using direct key
input.
- Intuitive control using cursor keys.
- Using item pages to set display
preferences.
- A variety of display formats.
- 50ms data; updating intervals,
compensation functions.
- A variety of integration functions:
User-defined function, Efficiency Calculation Function.
Direct Key Input Range
Settings |
|
The range indicator on
the WT3000 is a seven-segment green LED, so the set range can be
monitored at all times. The range can easily be switched using
the up and down arrows. |
 |
Easy Cursor Navigation and
Numerical Settings
|
|
Cursor keys can be used
to move the screen cursor in four different directions, so it is
easy and intuitive to set scaling factor and other settings. |
 |
Setting Display Preferences
|
Item pages
make it easy to set data you want to view for each
|
|
 |
The WT3000 has nine item pages for
displaying measurement values. Once you set the measurement
parameters you want displayed on a particular item page, you can
easily switch between entire groups of displayed parameters. For
example, the following settings make it easy to switch and
compare data.Page 1. Voltage,
current, active power, and current frequency for input element
1.
Page 2. Voltage, current, active power, and current frequency
for input element 2.
Page 3. Voltages for input elements 1, 2, 3 and 4.
Page 4. Currents for input elements 1, 2, 3 and 4.
Page 5. Power for input elements 1, 2, 3 and 4. |
Formats for Viewing Waveforms and Numerical
Values
|
|
The WT3000 lets you
display input signal waveforms in addition to numerical data.
This means you don't need to connect a special waveform analyzer
just to check signal waveforms.
In addition, the
optional advanced calculation function lets you display vectors
and bar graphs for enhanced visual presentation. The information
display shows voltage range, current range, filter, and scaling
value all together, making it easy to check your settings.
-
Waveforms up to
approximately 10kHz can be displayed accurately.
-
Models 760302, 760303
and 760304 require the optional advanced calculation function
(/G6).
|
 |
| |
Apparent Power Integration and Reactive Power Integrations
|
- Active power, current, apparent
power, reactive power
In addition to the active power integration function (WP) and
current integration function (q) included in earlier models,
the WT3000 also has a new apparent power integration function
(WS) and reactive power integration (WQ).
- A wide effective input range for
high-precision integration
The WT3000 has a wide
effective input range, from 1% to 130% of the measurement
range. This enables higher-precision integration measurements
on measurement subjects with current values that fluctuate
widely from large currents down to faint currents in the
standby state.
- Mean active power (using
user-defined settings)
Mean active power can be calculated over an integration
interval. This feature is useful for evaluating the power
consumed by intermittent-control instruments in which the
power value fluctuates.
|
 |
|
|